The Role And Principle Of Spo2 Sensor

The role and principle of Spo2 sensor

Most of the newly developed pulse oximeters use a clip sensor probe. 

When in use, the probe can be clamped on the fingertip. 

H6f

In addition, two light-emitting diodes placed side by side are fixed on the upper wall of the finger clamp, and the emission wavelengths are 660nm red light and 940nm infrared light respectively. 

hero-product-master

The lower wall is a photosensitive receiving device, 

which converts the red and infrared light transmitted through the finger into electrical signals. 

When the oximeter is running, the time-sharing driving circuit makes the two light-emitting diodes emit light at a certain time interval and with a relatively low duty cycle.

300

 Calculate the whole blood absorption rate a660 and a940, and then combine the experimental calibration coefficients A and B into the corresponding formula, 

to calculate the value of blood oxygen saturation.


And this is a continuous and non-invasive blood oxygen measurement method, 

so it will not cause any pain and side effects to people's bodies.