Home - Products - Pulse Oximeter - Details
CONTEC SPO2 Pulse Rate Patient Simulator Oxygen Saturation Simulator MS100
video
CONTEC SPO2 Pulse Rate Patient Simulator Oxygen Saturation Simulator MS100

CONTEC SPO2 Pulse Rate Patient Simulator Oxygen Saturation Simulator MS100


Packing:Individual Pack

Dimension:189(L)x 125 (W)x 56 (H) mm (including bracket)

Weigh: about 600 g (including a lithium battery)

Accessaries:A MS100 user manual、A 5.0V adaptor

Description

 

CONTEC SPO2 Pulse rate Patient Simulator Oxygen saturation Simulator MS100

 

product-700-50

MS100 SpO2 Simulator is a kind of Separated SpO2 simulator, small and light. It can perform a series of tests

for the oximeter by simulation means, and gives cognizance of veracity about the oximeter.

20240706152117

20240706151218

20240706151227

product-700-50

  • Separated connection between simulator probe and host, operation and test are more convenient.
  •  262K color and 320*240 TFT, adjustable brightness level.
  •  Film key-press, makes the operation more comfortable.
  •  Power with chargeable lithium battery, can display the charge information of battery.
  •  Key volume, can open or close.

 

Main Functions
◆ Oxygen saturation simulation;
◆ Pulse rate simulation;
◆ Patient state simulation preset;
◆ Can test the reaction time of device;
◆ Can simulate the SpO2 and pulse rate under different amplitude;
◆ Can test the performance of device under different interference source;
◆ Can select different R-curve.

 

Main Parameters

SpO2 1.Measuring range: 35%~100%
2.Resolution:1%
3.Error: When SpO2 is 75%~100%, the error is the larger between ±2% and ±P(P denotes the precision of the tested oximeter); when SpO2 is 74%~50%, the error is the larger between ±2% and ±P; no define when SpO2 is less than 50%.
Pulse rate 1.Measuring range: 25bpm~250bpm
2.Resolution: 5bpm
3.Error: 1%± 1bpm
Amplitude 1.Measuring range:0.000%~20.000%
2.Resolution:1% for 1.000%~20.000%; 0.1% for 0.100%~0.900%; 0.025% for 0.000%~0.075%.
Patient state 24 conditions presetting and 8 conditions as default after start. The number of condition can be adjusted by Motion level setup.
Voltage DC 3.6V~4.2V
Power Supply Voltage 3.7 rechargeable lithium battery × 1
Battery working life Charge and discharge no less than 500 times

 